Appendix B: Calibration
Model 2651A High Power System SourceMeter® Instrument Reference Manual
B-26
2651A-901-01 Rev. A / March 2011
22. If this step is not on the CALA sense mode, send the measure calibration command using the
multimeter and Model 2651A readings, and range setting for the parameters:
smua.measure.calibratev(
-range, Z_rdg, DMM_Z_rdg, FS_rdg, DMM_FS_rdg
)
Where:
-range
= The negative of the present calibration range
Z_rdg
= The zero Model 2651A measurement
DMM_Z_rdg
= The zero DMM measurement
FS_rdg
= The FS Model 2651A measurement
DMM_FS_rdg
= The FS DMM measurement
Typical Model 2651A 1 V range values:
smua.measure.calibratev(-1, -1e-4, -1e-6, -0.89, -0.896)
D. Be sure to complete each of the 22 steps of C for all 11 voltage source and measure function
steps in the
(on page B-19) before continuing to the current
calibration.
E. Select automatic polarity mode:
smua.cal.polarity = smua.CAL_AUTO
Step 3. Current calibration adjustment
Figure 116: Connections for current calibration (100 mA range and below)
LO
HI
Input
LO
Model 2651A
!
LINE RATING
100-240 VAC
50, 60 Hz
500VA MAX.
LINE FUSE
SLOWBLOW
5.0 A, 250 V
NO INTERNAL OPERATOR SERVICEABLE PARTS,
SERVICE BY QUALIFIED PERSONNEL ONLY.
REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
CHANNEL A
RS-232
DIGITAL I/O
LAN
!
IEEE-488
G
G G G
S
LO
SENSE/GUARD
S
HI
LO
HI
OUTPUT
40V, 50A MAX.
!
Amps
Digital multimeter