Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-46
2651A-901-01 Rev. A / March 2011
digio.trigger[N].mode
This attribute sets the mode in which the trigger event detector and the output trigger generator operate on the
given trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Digital I/O trigger N reset
Not saved
digio.TRIG_BYPASS
Usage
triggerMode
= digio.trigger[
N
].mode
digio.trigger[
N
].mode =
triggerMode
triggerMode
The trigger mode; see
Details
for values
N
Digital I/O trigger line (1 to 14)
Details
Set
triggerMode
to one of the following values:
Trigger mode values
triggerMode
Description
digio.TRIG_BYPASS or 0
Allows direct control of the line.
digio.TRIG_FALLING or 1
Detects falling-edge triggers as input; asserts a TTL-low pulse for
output.
digio.TRIG_RISING or 2
If the programmed state of the line is high, the
digio.TRIG_RISING
mode behaves similar to
digio.TRIG_RISINGA
. If the programmed state of the line is low,
the
digio.TRIG_RISING
mode behaves similar to
digio.TRIG_RISINGM
. This setting should only be used if
necessary for compatibility with other Keithley Instruments products.
digio.TRIG_EITHER or 3
Detects rising- or falling-edge triggers as input. Asserts a TTL-low
pulse for output.
digio.TRIG_SYNCHRONOUSA or 4
Detects the falling-edge input triggers and automatically latches and
drives the trigger line low. Asserting the output trigger releases the
latched line.
digio.TRIG_SYNCHRONOUS or 5
Detects the falling-edge input triggers and automatically latches and
drives the trigger line low. Asserts a TTL-low pulse as an output
trigger.
digio.TRIG_SYNCHRONOUSM or 6
Detects rising-edge triggers as input. Asserts a TTL-low pulse for
output.
digio.TRIG_RISINGA or 7
Detects rising-edge triggers as input. Asserts a TTL-low pulse for
output.
digio.TRIG_RISINGM or 8
Asserts a TTL-high pulse for output. Input edge detection is not
possible in this mode.
When programmed to any mode except
digio.TRIG_BYPASS
, the output state of the I/O line is controlled by
the trigger logic, and the user-specified output state of the line is ignored.
Use of either
digio.TRIG_SYNCHRONOUSA
or
digio.TRIG_SYNCHRONOUSM
is preferred over
digio.TRIG_SYNCHRONOUS
, as
digio.TRIG_SYNCHRONOUS
is provided for compatibility with the digital I/O
and TSP-Link triggering on other Keithley Instruments products.
To control the line state, set the mode to
digio.TRIG_BYPASS
and use the
digio.writebit()
and
digio.writeport()
commands.