Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-188
2651A-901-01 Rev. A / March 2011
smuX.contact.speed
This attribute stores the speed setting for contact check measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Not saved
smuX.CONTACT_FAST
Usage
speedSetting
= smu
X
.contact.speed
smu
X
.contact.speed =
speedSetting
speedSetting
The speed setting. Set to one of the following:
0
or
smu
X
.CONTACT_FAST
1
or
smu
X
.CONTACT_MEDIUM
2
or
smu
X
.CONTACT_SLOW
X
SMU channel (for example,
smua.contact.speed
applies to SMU Channel A).
Details
This setting controls the aperture of measurements made for contact check. It does not affect the
smu
X
.measure.nplc
aperture setting.
The speed setting can have a dramatic effect on the accuracy of the measurement (see specifications).
Example
smua.contact.speed = smua.CONTACT_SLOW
Configure contact check for higher accuracy
on SMU Channel A.
Also see
(on page 7-186)
(on page 7-209)
(on page 2-45)
(on page 2-37)