smuX.measure.delay
This attribute controls the measurement delay.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Not saved
smuX.DELAY_OFF
Usage
mDelay
= smu
X
.measure.delay
smu
X
.measure.delay =
mDelay
mDelay
Set to your desired measurement delay value (for example, to specify an additional
10 ms measurement delay, set the value to
0.010
)
You can also set it one of the following values:
0
or
smu
X
.DELAY_OFF
: No delay
-1
or
smu
X
.DELAY_AUTO
: Automatic delay value
X
Source-measure unit (SMU) channel (for example,
smua.measure.delay
applies
to SMU channel A).
Details
This attribute allows for additional delay (settling time) before taking a measurement. You can set
mDelay
to
smu
X
.DELAY_OFF
,
smu
X
.DELAY_AUTO
, or to a user-defined value (in seconds). A user-defined value will set
the delay used, regardless of range.
The
smu
X
.DELAY_AUTO
setting also causes a current range-dependent delay to be inserted when a current
measurement is requested. This happens when a current measurement command is executed, when the
measure action is being performed in a sweep, or after changing ranges during an autoranged measurement.
If
smu
X
.measure.count
is greater than 1, the measurement delay is only inserted before the first
measurement.
Example
smua.measure.delay = 0.010
Sets a 10 ms measurement delay for SMU channel A.
Also see
(on page 2-66)
(on page 7-194)
(on page 7-196)
(on page 7-209)