Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-216
2651A-901-01 Rev. A / March 2011
smuX.source.levelY
This attribute sets the source level.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall
Saved setups
0
Usage
sourceLevel
= smu
X
.source.level
Y
smu
X
.source.level
Y
=
sourceLevel
sourceLevel
The source value. Set to one of the following values: Voltage from 0 V to ±40 V.
Current from 0 A to ±20 A.
X
SMU channel (for example,
smua.source.levelv
applies to SMU Channel A).
Y
SMU source function (
v
= voltage,
i
= current).
Details
This attribute configures the output level of the voltage or current source.
If the source is configured as a voltage source and the output is on, the new
smu
X
.source.levelv
setting will
be sourced immediately. If the output is off or the source is configured as a current source, the voltage level will
be sourced when the source is configured as a voltage source and the output is turned on.
If the source is configured as a current source and the output is on, the new
smu
X
.source.leveli
setting will
be sourced immediately. If the output is off or the source is configured as a voltage source, the current level will
be sourced when the source is configured as a current source and the output is turned on.
The sign of
sourceLevel
dictates the polarity of the source. Positive values generate positive voltage or current
from the high terminal of the source relative to the low terminal. Negative values generate negative voltage or
current from the high terminal of the source relative to the low terminal.
The
reset()
function sets the source levels to 0 V and 0 A.
Example
smua.source.levelv = 1
Sets voltage source of SMU
Channel A to 1 V.
Also see
(on page 4-2)