Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-220
2651A-901-01 Rev. A / March 2011
Example
smua.source.offlimiti = 10e-3
Changes the normal output-off
mode limit to 10 mA for SMU
channel A.
Also see
(on page 7-220)
smuX.source.offmode
This attribute sets the source output-off mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall
Saved setups
smuX.OUTPUT_NORMAL
Usage
sourceOffMode
= smu
X
.source.offmode
smu
X
.source.offmode =
sourceOffMode
sourceOffMode
The output-off setting; set to one of the following values:
0
or
smu
X
.OUTPUT_NORMAL
: Configures the source function according to
smu
X
.source.offfunc
attribute.
1
or
smu
X
.OUTPUT_ZERO
: Configures source to output 0 V as
smu
X
.OUTPUT_NORMAL
but has different compliance handling (see the
Details
below)
2
or
smu
X
.OUTPUT_HIGH_Z
: Opens the output relay when the output is turned off
X
Source-measure unit (SMU) channel (for example,
smua.source.offmode
applies to SMU channel A)