smuX.trigger.count
This attribute sets the trigger count in the trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Not saved
1
Usage
triggerCount
= smu
X
.trigger.count
smu
X
.trigger.count =
triggerCount
triggerCount
The trigger count is the number of times the SMU will iterate in the trigger layer for
any given sweep.
X
SMU channel (for example,
smua.trigger.count
applies to SMU Channel A).
Details
During a sweep, the SMU will iterate through the trigger layer of the trigger model this many times. After
performing this many iterations, SMU will return to the arm layer.
If this count is set to zero (
0
), the SMU will stay in the trigger model indefinitely until aborted.
Also see
(on page 3-32)
smuX.trigger.endpulse.action
This attribute enables or disables pulse mode sweeps.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Not saved
smuX.SOURCE_HOLD
Usage
pulseAction
= smu
X
.trigger.endpulse.action
smu
X
.trigger.endpulse.action =
pulseAction
pulseAction
The pulse mode setting; set to one of the following values (see details for definition):
1
or
smu
X
.SOURCE_HOLD
0
or
smu
X
.SOURCE_IDLE
X
SMU channel (for example,
smua.trigger.endpulse.action
applies to SMU
channel A)
Details
When set to
smu
X
.SOURCE_HOLD
, this attribute disables pulse mode sweep, holding the source level for
remainder of step. When set to
smu
X
.SOURCE_IDLE
, this attribute enables pulse mode sweep, setting the
source level to the programmed (idle) level at the end of the pulse.