Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-255
Bit Value
Description
B0
Not used
Not applicable.
B1
status.measurement.buffer_available.SMUA
Set bit indicates that there is at least one
reading stored in either or both of the dedicated
reading buffers.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event buffer available summary enable register, set
status.measurement.buffer_available.enable =
status.measurement.buffer_available.SMUA
.
In addition to the above constant,
measurementRegister
can be set to the decimal equivalent of the bit to set.
Example
status.measurement.buffer_available.enable =
status.measurement.buffer_available.SMUA
Sets the SMUA bit of the
measurement event buffer available
summary enable register.
Also see
(on page E-23)
status.measurement.current_limit.*
This attribute contains the measurement event current limit summary registers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
2 (All bits set)
Usage
measurementRegister
= status.measurement.current_limit.condition
measurementRegister
= status.measurement.current_limit.enable
measurementRegister
= status.measurement.current_limit.event
measurementRegister
= status.measurement.current_limit.ntr
measurementRegister
= status.measurement.current_limit.ptr
status.measurement.current_limit.enable =
measurementRegister
status.measurement.current_limit.ntr =
measurementRegister
status.measurement.current_limit.ptr =
measurementRegister
measurementRegister
The measurement event current limit summary register's status. A zero (0)
indicates no bits set (also send 0 to clear all bits); the only valid value other
than 0 is 2