Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-266
2651A-901-01 Rev. A / March 2011
status.operation.*
These attributes manage the status model's operation status register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
31,769 (All bits set)
Usage
operationRegister
= status.operation.condition
operationRegister
= status.operation.enable
operationRegister
= status.operation.event
operationRegister
= status.operation.ntr
operationRegister
= status.operation.ptr
status.operation.enable = o
perationRegister
status.operation.ntr = o
perationRegister
status.operation.ptr = o
perationRegister
operationRegister
The operation status register's status. A zero (0) indicates no bits set (also send 0 to
clear all bits); other values indicate various bit settings
Details
These attributes are used to read or write to the operation status registers. Reading a status register returns a
value. The binary equivalent of the returned value indicates which register bits are set. The least significant bit of
the binary number is bit b0, and the most significant bit is Bit B15. For example, if a value of
2.04
(which is 20,480) is read as the value of the condition register, the binary equivalent is 0101 0000 0000 0000.
This value indicates that bit B12 and bit B14 are set.
B15 B14 B13 B12 B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0
** > > > > > > > > > > > > > > *
0
1
0
1
0
0
0
0
0
0
0
0
0
0
0
0
* Least significant bit
** Most significant bit
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on page E-
Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings: