Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-268
2651A-901-01 Rev. A / March 2011
Example 2
-- decimal 20480 = binary 0101 0000 0000 0000
operationRegister = 20480
status.operation.enable = operationRegister
Sets the USER and PROG bits of the
operation status enable register using a
decimal value.
Also see
status.operation.calibrating.*
This attribute contains the operation status calibration summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
2 (All bits set)
Usage
operationRegister
= status.operation.calibrating.condition
operationRegister
= status.operation.calibrating.enable
operationRegister
= status.operation.calibrating.event
operationRegister
= status.operation.calibrating.ntr
operationRegister
= status.operation.calibrating.ptr
status.operation.calibrating.enable =
operationRegister
status.operation.calibrating.ntr =
operationRegister
status.operation.calibrating.ptr =
operationRegister
operationRegister
The operation calibrating event register's status. A zero (0) indicates no bits set
(also send 0 to clear all bits); the only valid value other than 0 is 2
Details
These attributes are used to read or write to the operation status calibration summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on page E-
Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings:
Bit Value
Description
B0
Not used
Not applicable.
B1
status.operation.calibrating.SMUA
Set bit indicates that SMU A is unlocked for
calibration.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2 to
B15
Not used
Not applicable.