Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-341
Also see
trigger.timer[N].delaylist
This attribute sets an array of timer intervals that are used when triggered.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Trigger timer N reset
Not saved
{10e-6}
Usage
intervals
= trigger.timer[
N
].delaylist
trigger.timer[
N
].delaylist =
intervals
intervals
Table of delay intervals in seconds
N
Trigger timer number (1 to 4)
Details
Each time the timer is triggered, it uses the next delay period from the array.
After all elements in the array have been used, the delays restart at the beginning of the list.
If the array contains more than one element, the average of the delay intervals in the list must be
≥
50 µs.
Example
trigger.timer[3].delaylist = {50e-6, 100e-6, 150e-6}
DelayList = trigger.timer[3].delaylist
for x = 1, table.getn(DelayList) do
print(DelayList[x])
end
Set a delay list on trigger timer 3
with three delays (50 µs, 100 µs,
and 150 µs).
Read the delay list on trigger timer
3:
Output (assuming the list was set
with three delays (50 µs, 100 µs,
and 150 µs):
5.000000000e-05
1.000000000e-04
1.500000000e-04
Also see
trigger.timer[N].EVENT_ID
This constant specifies the trigger timer event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant Yes