Model 2651A High Power System SourceMeter® Instrument Reference Manual
Appendix B: Calibration
2651A-901-01 Rev. A / March 2011
B-13
Voltage source accuracy
Follow the steps below to verify that the Model 2651A output voltage accuracy is within specified
limits. This test involves setting the output voltage to each full-range value and measuring the
voltages with a precision digital multimeter.
1. With the power off, connect the digital multimeter (DMM) to the Model 2651A output terminals
using 4-wire connections, as shown below.
Figure 114: Connections for voltage verification
LO
S
LO
S
HI
HI
Input
LO
Model 2651A
Digital multimeter
!
LINE RATING
100-240 VAC
50, 60 Hz
500VA MAX.
LINE FUSE
SLOWBLOW
5.0 A, 250 V
NO INTERNAL OPERATOR SERVICEABLE PARTS,
SERVICE BY QUALIFIED PERSONNEL ONLY.
REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
CHANNEL A
RS-232
DIGITAL I/O
LAN
!
IEEE-488
G
G G G
S
LO
SENSE/GUARD
S
HI
LO
HI
OUTPUT
40V, 50A MAX.
!
Input HI
2. Select the multimeter DC volts measuring function.
3. Press
the
SRC
key to source voltage and make sure the source output is turned on.
4. Enable the Model 2651A 4-wire (remote sense) mode by pressing the
CONFIG
key and then the
SRC
key.
5. Select
V-SOURCE > SENSE-MODE > 4-WIRE
.