Section 2: General operation
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2-56
2651A-901-01 Rev. A / March 2011
Using shielding and guarding together
The following figures show connections for a test system that uses a noise shield, a safety shield, and
guarding. The guard shields are connected to the driven guard (G) of the SMU. The noise shield is
connected to SMU LO. The safety shield is connected to the chassis and to a safety earth ground.
Figure 22: Connections for noise shield, safety shield, and guarding
G
G G G
S
LO
SENSE/GUARD
S
HI
LO
HI
OUTPUT
40V, 50A MAX.
Chassis
ground
Safety
earth
ground
Metal guard
shield
HI
G
LO
HI
LO
Chassis
Metal safety shield
Metal noise shield
Metal guard shield
Model 2651A chassis
connected to chassis
earth ground through the
power cord
Test
circuit