Section 2: General operation
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2-72
2651A-901-01 Rev. A / March 2011
Setting speed
Speed is set from the SPEED configuration menu and is structured as follows.
Front-panel speed configuration
Press
SPEED
(or use the CONFIG menu) to display the following menu items:
•
FAST:
Selects the fast analog-to-digital converter (ADC)
•
MED:
Sets the speed to 0.10 PLC
•
NORMAL:
Sets speed to 1.00 PLC
•
HI-ACCURACY:
Sets speed to 10.00 PLC
•
OTHER:
Used to set speed to any PLC value from 0.001 to 25
NOTE
Selecting the FAST menu item selects the fast analog to digital converter (ADC). All other menu items
on the speed menu use the integrating ADC.
NOTE
The SPEED setting affects all measurement functions. After setting speed, display resolution can be
changed using the DIGITS key.
Remote speed programming
Speed command
The following table summarizes commands to control speed. See
for more information.
Speed command
Command Description
smua.measure.nplc = nplc
Sets the speed of the integrating ADC only (nplc =
0.001 to 25). *
smua.measure.adc = adc
Select analog to digital converter (adc =
smua.ADC_INTEGRATE
or
smua.ADC_FAST
).
* The speed setting is global and affects all measurement functions.
Speed programming example
Use the NPLC command to set the speed. The programming example below illustrates how to set the
speed to 10 PLC:
-- Set NPLC to 10.
smua.measure.nplc = 10