Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2657A-901-01 Rev. B/December 2012
3-19
Dynamically allocated buffer example
The programming example below illustrates how to store data to an allocated buffer called
mybuffer
. The Model 2657A stores 100 current readings in
mybuffer
and then recalls all the
readings.
-- Restore Model 2657A defaults.
smua.reset()
-- Select measure I autorange.
smua.measure.autorangei = smua.AUTORANGE_ON
-- Select measure V autorange.
smua.measure.autorangev = smua.AUTORANGE_ON
-- Select ASCII data format.
format.data = format.ASCII
-- Set buffer count to 100.
smua.measure.count = 100
-- Set measure interval to 0.1 s.
smua.measure.interval = 0.1
-- Select source voltage function.
smua.source.func = smua.OUTPUT_DCVOLTS
-- Output 100 V.
smua.source.levelv = 100
-- Turn on output.
smua.source.output = smua.OUTPUT_ON
-- Create a temporary reading buffer.
mybuffer = smua.makebuffer(smua.measure.count)
-- Store current readings in mybuffer.
smua.measure.overlappedi(mybuffer)
-- Wait for buffer to fill.
waitcomplete()
-- Turn off output.
smua.source.output = smua.OUTPUT_OFF
-- Output readings 1 to 100 from mybuffer.
printbuffer(1, 100, mybuffer)
-- Delete mybuffer.
mybuffer = nil
Sweep operation
Overview
The Model 2657A can generate DC and pulsed sweeps to perform source-only sweeps,
source-and-measure sweeps, or measure-only sweeps. There are three sweep types shown in the
following figure: DC and pulsed linear staircase sweeps (A), DC and pulsed logarithmic staircase
sweeps (B), and DC and pulsed list sweeps (C). Details about each kind of sweep follow the figure.