Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-42
2657A-901-01 Rev. B/December 2012
This function does not cause the specified
smu
to output a pulse train. It does check to see if all the pulse
dimensions can be achieved, and if they can, assigns the indicated
tag
or index to the pulse train.
Figure 113: ConfigPulseVMeasureISweepLin()
Example
ConfigPulseVMeasureISweepLin(smua, 0,
100, 1000, 1e-3, 10e-3, 50e-3, 16,
smua.nvbuffer1, 4)
Set up a pulsed sweep that uses System
SourceMeter
®
instrument channel A. The pulsed
sweep starts at 100 V, ends at 1000 V, and returns to
a 0 V bias level between pulses. Each pulsed step is
on for 10 ms, and then at the bias level for 50 ms.
The current limit is 1 mA during the entire pulsed
sweep. The pulse train is comprised of 16 pulsed
steps, and the pulse train is assigned a
tag
index of
4.
Also see
ConfigPulseVMeasureISweepLog()
(on page 5-20) function configures a voltage pulse train with a current measurement at
each point.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Measurement (1/NPLC seconds)
bias
step
step
ton
toff
start
stop
Pulse initiated here
Last source value before pulse train was initiated