Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-160
2657A-901-01 Rev. B/December 2012
script.new()
This function creates a script.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Usage
scriptVar
= script.new(
code
)
scriptVar
= script.new(
code
,
name
)
scriptVar
The name of the variable that will reference the script
code
A string containing the body of the script
name
The name of the script
Details
The
name
parameter is the name that is added to the
script.user.scripts
table. If
name
is not given, an
empty string will be used, and the script will be unnamed. If the name already exists in
script.user.scripts
,
the existing script's
name
attribute is set to an empty string before it is replaced by the new script.
Note that
name
is the value that is used for the instrument front panel display. If this value is not defined, the
script will not be available from the instrument front panel.
You must save the new script into nonvolatile memory to keep it when the instrument is turned off.
Example 1
myTest8 = script.new(
"display.clear() display.settext('Hello from myTest8')", "myTest8")
myTest8()
Creates a new script referenced by the variable
myTest8
with the name "
myTest8"
.
Runs the script. The instrument displays "
Hello from myTest8
".
Example 2
autoexec = script.new(
"display.clear() display.settext('Hello from autoexec')", 'autoexec')
Creates a new autoexec script that clears the display when the instrument is turned on and displays
"
Hello from autoexec
".
Also see
Create a script using the script.new() command
Global variables and the script.user.scripts table
(on page 6-38)
(on page 6-4)
(on page 7-167)
(on page 7-161)
Summary of Contents for 2657A