Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-258
2657A-901-01 Rev. B/December 2012
Usage
measurementRegister
= status.measurement.overvoltage.condition
measurementRegister
= status.measurement.overvoltage.enable
measurementRegister
= status.measurement.overvoltage.event
measurementRegister
= status.measurement.overvoltage.ntr
measurementRegister
= status.measurement.overvoltage.ptr
status.measurement.overvoltage.enable =
measurementRegister
status.measurement.overvoltage.ntr =
measurementRegister
status.measurement.overvoltage.ptr =
measurementRegister
measurementRegister
The status of the measurement overvoltage summary register; a zero (0)
indicates no bits set (also send 0 to clear all bits); the only valid value other
than 0 is 2
Details
These attributes are used to read or write to the measurement event overvoltage summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is bit B0, and the most significant bit is bit B15. For example,
assume the value 2 is returned for the enable register. The binary equivalent is 0000 0000 0000 0010. This value
indicates that bit B1 (SMUA) is set.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on
Enable and transition registers
(on page E-19). The individual bits of this register are defined in
the following table.
Bit Value
Description
B0
Not used
Not applicable.
B1
status.measurement.overvoltage.SMUA
Set bit indicates that an overvoltage condition
has been detected.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2-B15
Not used
Not applicable.
As an example, to set bit B1 of the measurement event overvoltage summary enable register, set
status.measurement.overvoltage.enable = status.measurement.overvoltage.SMUA
.
In addition to the above constants,
measurementRegister
can be set to the numeric equivalent of the bit to
set. For example, to set bit B1, set
measurementRegister
to 2.
Example
status.measurement.overvoltage.enable =
status.measurement.overvoltage.SMUA
Sets the SMU A bit of the
measurement overvoltage summary
enable register using a constant.
Also see
Summary of Contents for 2657A
Page 3: ......