Model 2657A High Power System SourceMeter® Instrument Reference Manual
Appendix B: Calibration
2657A-901-01 Rev. B/December 2012
B-11
Figure 128: Alternate connections for voltage verification (sample voltage divider shown)
10 M
Ω
200 k
Ω
HI
Sense HI
Sense LO
LO
10 M
Ω
10 M
Ω
10 M
Ω
10 M
Ω
10 M
Ω
LINE RATING
100-240 VAC
50, 60 Hz
350 VA MAX.
LINE FUSE
SLOWBLOW
5.0 A, 250 V
NO INTERNAL OPERATOR SERVICEABLE PARTS,
SERVICE BY QUALIFIED PERSONNEL ONLY.
REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
INPUT/OUTPUT CURRENT: 120 mA MAX.
RS-232
DIGITAL I/O
LAN
!
IEEE-488
!
!
SENSE
LO
HI
LO
GUARD
SENSE
HI
LO
3030 VDC
MAX.
250 V
MAX.
* This sample voltage divider should
be used when verifying/calibrating
the 1500 V and 3000 V ranges with
a meter that can measure < 1000 V.
Use a direct connection for 200 V
and 500 V ranges.
Model 2657A
Digital multimeter
Test fixture
Input
HI
Input
LO
Sample voltage divider*
Pin 22 = +5 V
Pin 24 = Interlock (INT)
Summary of Contents for 2657A
Page 3: ......