Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 2: General operation
2657A-901-01 Rev. B/December 2012
2-69
To configure the output-off function from the front panel:
1. Press
the
CONFIG
key.
2. Press
the
OUTPUT ON/OFF
control. The CONFIGURE OUTPUT A menu is displayed.
3. In the CONFIGURE OUTPUT A menu, select
OFF-STATE
to display the OUTPUT OFF STATE
A menu.
4. With the OUTPUT OFF STATE A menu displayed, select
FUNCTION
to display the OFF
FUNCTION A menu.
5. In the OFF FUNCTION A menu, select
CURRENT
or
VOLTAGE
.
To configure the output-off function remotely:
To set 0 V output with current limit set by the
smua.source.offlimiti
attribute:
smua.source.offfunc = smua.OUTPUT_DCVOLTS
To set 0 A output with voltage limit set by the
smua.source.offlimitv
attribute:
smua.source.offfunc = smua.OUTPUT_DCAMPS
Output-off limits (compliance)
You can set output-off limits (compliance) for the current and voltage output-off functions using the
CONFIG menu on the Model 2657A front panel, or by setting the
smua.source.offlimit
Y
attribute from a remote interface. The output-off limits only apply when the output-off mode is normal.
Setting the output-off limit for CURRENT (
smua.source.offlimiti
) specifies the current limit for
the voltage source; setting the output-off limit for VOLTAGE (
smua.source.offlimitv
) specifies
the voltage limit for the current source.
Setting output-off limits
To configure output-off limits from the front panel:
1. Press
the
CONFIG
key.
2. Press
the
OUTPUT ON/OFF
control. The CONFIGURE OUTPUT A menu is displayed.
3. In the CONFIGURE OUTPUT A menu, select
OFF-STATE
to display the OUTPUT OFF STATE
A menu.
4. With the OUTPUT OFF STATE A menu displayed, select
LIMIT
to display the OFF LIMIT A
menu.
5. In the OFF LIMIT A menu, select
CURRENT
or
VOLTAGE
.
6. Use the left or right arrow keys or turn the navigation wheel
to select the current or voltage
limit, and then press the
ENTER
key or the navigation wheel
to save your settings.
7. Press
the
EXIT
key to return to the main display.
To set the current limit in NORMAL output-off mode remotely:
smua.source.offlimiti = iValue
To set the voltage limit in NORMAL output-off mode remotely:
smua.source.offlimitv = vValue