3-32
Basic DMM Operation
Model 2701 User’s Manual
10M
Ω
and 100M
Ω
ranges – Open sense lead detection for the 10M
Ω
and 100M
Ω
detection
is slightly different and is shown in Figure 3-13. Detection is performed at Sense Lo only.
Sense Hi is not used. It does not need to be connected to the DUT. When the Sense Lo lead
opens, the Sense Lo terminal will drift to -15mV and trip the “OVRFLOW” message.
Figure 3-13
Open Sense Lo lead detection (10M
Ω
and 100M
Ω
ranges)
4-wire common-side (CSID) ohms measurements (7701 module)
For normal 4-wire ohms measurements using a switching module, channels are paired to
provide the switch paths for input and sense. Each tested DUT requires two input
channels. For example, the 7700 module has 20 channels. With the
Ω
4 function selected,
channel 1 is paired to channel 11, channel 2 is paired to channel 12, and so on. This
configuration allows up to 10 DUT to be tested.
The 7701 module has 32 input channels. For normal 4-wire ohms measurements, up to 16
DUT can be tested. However, this module can be configured for common-side (CSID)
4-wire ohms measurements, allowing up to 32 DUT to be tested.
With a 7701 module installed, the 4-wire ohms mode can be selected using the following
key-press sequence:
1.
Press SHIFT and then press CARD.
2.
Select CONFIG.
3.
Select slot that has the 7701 (i.e., SLOT1: 7701).
4.
Select 4W MODE: NORM (normal) or CSID (common-side).
Sense HI
Input HI
Input Lo
Sense Lo
S/W Detection
I-Source
10M
Ω
DUT
1mA
-15mV
Open Sense Lo
Lead
OVRFLW
M
Ω
2701 Reading
(10M
Ω
range)
2701
H/W Detection
S/W Detection
Not Used*
Since Sense Hi is not used for the measurement, there is no open
test lead detection for Sense Hi. The Sense Hi test lead does not
need to be connected to the DUT.
*
2701-900-01.book Page 32 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...