3-46
Basic DMM Operation
Model 2701 User’s Manual
Frequency and period measurements
The Model 2701 can make frequency measurements from 3Hz to 500kHz on voltage
ranges of 100mV, 1V, 10V, 100V, and 750V. Period (1 / frequency) measurements can be
taken from 2µs to 333ms on the same voltage ranges as the frequency.
Input impedance:1M
Ω
|| <100pF, AC coupled.
The instrument uses the volts input to measure frequency. The AC voltage range can be
changed with the RANGE and keys. The signal voltage must be greater than 10% of
the full-scale range.
CAUTION
The voltage limit is subject to the 8
×
10
7
VHz product.
Trigger level
Frequency and period use a zero-crossing trigger, meaning that a count is taken when the
frequency crosses the zero level. The Model 2701 uses a reciprocal counting technique to
measure frequency and period. This method generates constant measurement resolution
for any input frequency. The multimeter’s AC voltage measurement section performs
input signal conditioning.
Gate time
The gate time is the amount of time the Model 2701 uses to sample frequency or period
readings. Use the RATE key to set the gate time; SLOW sets the gate time to 1.0 sec, MED
sets it to 0.1 sec, and FAST sets it to 0.01 sec. For remote programming, the gate time can
be set from 0.01 to 1.0 sec using the FREQuency:APERture and PERiod:APERture
commands (Table 3-7). Note however, that if you set a gate time other than 1.0, 0.1, or
0.01 sec, the SLOW, MED, and FAST annunciators will be off.
The Model 2701 completes a reading when it receives its first zero-crossing after the gate
time expires. In other words, the reading is completed 1/2 cycle after the gate time has
expired. For example, to sample a 3Hz frequency, you may wait up to three seconds before
the Model 2701 returns a reading.
2701-900-01.book Page 46 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...