Model 2701 User’s Manual
Basic DMM Operation
3-51
Remote programming for basic measurements
Basic measurement commands
NOTE
When measurements are performed, the readings are fed to other enabled
processing operations. Appendix D explains “Data flow (remote operation)”
and the commands used to return the various processed readings.
Commands to perform basic measurements are listed in Table 3-7.
Table 3-7
Basic measurement commands
Commands
1
Description
Default Ref
Select measurement function
[SENSe[1]]
Optional root command.
:FUNCtion <name> [, <clist>]
Select measurement function:
<name> = ‘VOLTage[:DC]’,
‘VOLTage:AC’, ‘CURRent[:DC]’,
‘CURRent:AC’, ‘RESistance’,
‘FRESistance’, ‘CONTinuity’,
‘FREQuency’, or ‘PERiod’.
Note: [:DC] is optional.
VOLT
a
DCV function
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:IDIVider <b>
Enable/disable DCV input divider;
<b> = ON or OFF.
OFF
Ω
4 function
[SENSe[1]]
Optional root command.
:FRESistance:OCOMpensated <b>
[, <clist>]
Enable/disable offset-compensated ohms. OFF
b
SYSTem
Root command for SYSTem subsystem.
:FRESistance:TYPEx, <name>
Select 4-wire ohms mode;
<name> = NORMal or CSIDe.
x = Slot number (1 or 2).
NORM
2701-900-01.book Page 51 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...