Model 2701 User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-9
mX+b rel
The mX+b function can be used to manually establish a rel value. To do this, set the scale
factor (m) to 1 and set the offset (b) to the rel value. Each subsequent reading will be the
difference between the actual input and the rel value (offset).
Percent
This math function determines percent deviation from a specified reference value. The
percent calculation is performed as follows:
where:
Input is the normal display reading.
Reference is the user entered constant.
Percent is the displayed result.
NOTE
When using Rel, the rel’ed reading of the input signal is used by the percent
calculation.
Percent configuration
1.
Press SHIFT and then MATH to display the math menu.
2.
Press the RANGE and key to display “PERCENT” and press ENTER to dis-
play the present reference value:
REF +1.000000
(factory default)
3.
Key in the reference value. The and keys control cursor position and the
and keys increment/decrement the digit value. To change range, place the cursor
on the multiplier and use the and keys (m =
×
0.001,
=
×
1, K =
×
1000, and
M =
×
1,000,000). With the cursor on the polarity sign, the and keys toggle
polarity.
4.
Press ENTER. The MATH annunciator will turn on and the result of the calculation
will be displayed. Note that the calculation will be applied to all measurement
functions.
5.
To disable mX+b, again press SHIFT and then MATH. The MATH annunciator
will turn off.
NOTE
The result of the percent calculation is positive when the input exceeds the
reference and negative when the input is less than the reference.
The result of the percent calculation may be displayed in exponential notation.
For example, a displayed reading of +2.500E+03% is equivalent to 2500%
(2.5K%).
Percent
Input
Reference
–
Reference
-------------------------------------------
100%
×
=
2701-900-01.book Page 9 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...