Model 2701 User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-11
Basic operation
NOTE
If using switching module inputs, make sure the front panel INPUTS switch is set
to the REAR position (in). If using the front panel inputs, the switch must be in
the FRONT position (out).
1.
Configure and enable the mX+b, percent, or reciprocal (1/X) math function as
previously explained.
2.
Select the desired measurement function.
3.
Apply the signal to be measured to a switching channel input or to the front panel
inputs.
NOTE
For the Model 7700 switching module, channels 21 and 22 are available for
DCI and ACI. Channels 1 through 20 are available for all other functions.
4.
If using a switching module (REAR inputs selected), use the or key to select
(close) the input channel. If using the front panel inputs (FRONT inputs selected),
it does not matter if a switching channel is closed. The result of the math
calculation will be displayed.
Scanning
When a simple scan is configured, the present math calculation will apply to all channels
in the scan. When an advanced scan is configured, each channel can have its own unique
math setup. Details to configure and run a scan are provided in Section 7.
For remote programming, the <clist> parameter is used to configure channels for a scan.
2701-900-01.book Page 11 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...