8-20
Triggering
Model 2701 User’s Manual
Reference
a.
ABORt
— With continuous initiation disabled, the 2701 goes into the idle state.
With continuous initiation enabled, operation continues at the top of the trigger
model.
b.
INITiate
— Whenever the instrument is operating within the trigger model,
sending this command causes an error and will be ignored.
c.
INITiate:CONTinuous <b>
— With continuous initiation enabled, you cannot
use the READ? command or set sample count (SAMPle:COUNt) greater than one.
d.
FETCh?
READ?
— See Section 3, Section 13, and Appendix D for details on using these
commands to trigger and retrieve readings.
NOTE
[SENSe[1]]:DATA[:LATest]? and [SENSe[1]]:DATA:FRESh? can be
used to retrieve the last reading. These commands are also explained in
Section 3, Section 13, and Appendix D.
e.
TRIGger:SOURce <name>
— With the timer control source selected, use the
TRIGger:TIMer command to set the interval.
f.
DELay:AUTO <b>
— The auto delay times are listed in Table 8-1. Disabling
auto delay sets the delay time to 0.
g.
TRIGger:SIGNal
— Send this action command to bypass the control source
when you do not wish to wait for the programmed event to occur. The instrument
must be waiting at the control source for the event when this command is sent.
Otherwise, an error occurs and the command is ignored.
h.
SAMPle:COUNt
— A sample count >1 specifies how many readings will
automatically be stored in the buffer. However, with continuous initiation enabled,
you cannot set the sample count greater than one.
i.
[SENSe[1]]:HOLD
commands — Hold cannot be set when the scanner is enabled
(ROUTe:SCAN:LSEL INT). Sending a hold command will result in a settings
conflict error (-221).
Programming example
The following program fragment triggers (and stores in the buffer) 10 readings:
*RST
' Restore *RST defaults.
TRAC:CLE
' Clear buffer.
TRIG:DEL 0.5
' Set delay for 0.5sec.
SAMP:COUN 10
' Set sample count to 10.
READ?
' Trigger, store, and request readings.
2701-900-01.book Page 20 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...