10-4
Remote Operations
Model 2701 User’s Manual
Password
A user-defined password can be used to disable protected commands. Most Model 2701
commands are protected. When the use of password is enabled, there are commands to
either disable or enable the protected commands.
At the factory, the Model 2701 is assigned the following password name:
“DEFAULT”
The password is case sensitive and must be enclosed in quotes, as shown above.
NOTE
From the front panel, the password can be cleared. When cleared, the use of
password is disabled and the password name resets to “DEFAULT”. The
password can be cleared using the SETUP menu as follows:
•
Press SHIFT
•
Press SETUP
•
Use the edit keys to display CLR PASSWORD: N or Y
where: N = no (do not clear) and Y = yes (clear)
•
Press ENTER
The password commands are listed in Table 10-1. The password commands are used to
perform the following operations:
•
Disable or enable the use of a password. Disabling the use of a password enables
all protected commands:
SYSTem:PASSword:ENABle
OFF
‘ Disable password.
SYSTem:PASSword:ENABle
ON ‘
Enable
password.
NOTE
The password can also be enabled or disabled from the Model 2701’s
internal web page (see “Internal web page,” page 10-16).
•
Disable protected commands. This places the Model 2701 in the
password-protected mode. In this mode, the only commands that can be sent are
the ones listed in Note 2 at the bottom of Table 10-1. Example using “DEFAULT”
as the password:
SYSTem:PASSword:CDISable
“DEFAULT”
NOTE
If the password is disabled, sending the :CDISable command
automatically enables the password and then places the Model 2701 in
the password-protected mode.
2701-900-01.book Page 4 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...