1-16
Getting Started
Model 2701 User’s Manual
Figure 1-3
Power module
Line frequency
The Model 2701 will operate at line frequencies from 45Hz to 66Hz and 360Hz to 440Hz.
There are no user-settings for line frequency. It is automatically sensed at power-up. The
following command can be used to read the line frequency:
SYSTem:LFRequency?
' Query power line frequency.
Setting line voltage and replacing fuse
A rear panel fuse located next to the AC receptacle protects the power line input of the
instrument. If the line voltage setting needs to be changed or the line fuse needs to be
replaced, perform the following steps.
WARNING
Make sure the instrument is disconnected from the AC line and other
equipment before changing the line voltage setting or replacing the
line fuse.
1.
Place the tip of a flat-blade screwdriver into the power module by the fuse holder
assembly (Figure 1-3). Gently push in and up. Release pressure on the assembly
and its internal spring will push it out of the power module.
Model 2701
KEITHLEY
SLOT COVER
TRIGGER
LINK
ETHERNET
10/100 BaseT
Link/Act
100bT
RS232
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
!
!
SLT
2
SLT
1
DIGITAL I/O
(EXT. TRIG.)
CAT I
Fuse
Spring
Window
Line
Voltage
Selector
Fuse Holder Assembly
120
240 220
100
~LINE RATING
50, 60, 400Hz
80VA MAX
250mAT
125mAT
220V
240V
100V
120V
FUSE
LINE
2701-900-01.book Page 16 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...