OComp (offset-compensated ohms)
The Model 2701 performs a normal ohms measurement by sourcing a known current (I),
measuring the voltage (V), and then calculating the resistance (R = V/I). Offset-
compensated ohms cancels the effects of thermal EMFs which can adversely affect low-
resistance measurements.
With OCOMP enabled, the Model 2701 performs one normal resistance measurement and
then loops back to perform a second resistance measurement with the internal current
source set to its lowest level. The offset-compensated ohms reading is then calculated as
shown in Figure D-2.
NOTE
For details on OCOMP measurements, see“Offset-compensated ohms,”
page 3-24.
Filter
The filter is used to stabilize noisy readings. With the filter enabled, the specified number
of readings are averaged to yield a single filtered reading. There are two types of filters:
moving and repeating.
A filter stack is used to temporarily store the specified number of readings to be averaged.
In general, for the moving filter, each measurement process adds a reading to the stack
(oldest reading discarded) and then averages the stack to yield a filtered reading. For the
repeating filter, each measurement process fills the stack with new readings (all previous
readings discarded) and then averages the stack to yield a filtered reading.
NOTE
For details on filter operation, see “Filter,” page 4-14.
Output trigger pulse (VMC)
An output trigger pulse from the Model 2701 can be used to trigger an external instrument
to perform an operation. In general, a trigger pulse is the output at this point in the
flowchart for each processed reading.
An exception is the SCAN function for scanning. For the SCAN function, an output
trigger is not output until after the specified number of channels (as set by the sample
counter) are scanned.
NOTE
For details on scan operation, see “Trigger models,” page 7-4.
2701-900-01.book Page 4 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...