2-14
Close/Open Switching Module Channels
Model 2701 User’s Manual
Remote programming example (system channel operation)
The following example assumes a Model 7700 installed in slot 1, and the
Ω
4 function of
the Model 2701 is selected. This command sequence connects channel 101 and its paired
channel (111) to DMM Input and Sense as shown in Figure 2-2.
ROUT:OPEN:ALL
' Open all channels.
ROUT:CLOS (@101)
' Close channels 101, 111, 123, 124, and 125.
Non-amp and non-measure switching modules
There are Keithley switching modules that do not support current measurements and there
are modules that do not support any measurements at all.
Non-amps module — With an amps function selected (DCI or ACI), system channel oper-
ation cannot be used to close channels on that module.
Non-measure module — For front panel operation, system channel operation cannot be
used to close channels. For remote programming, system channel operation can be used,
but only the one specified channel will close. All other channels on the module will open.
Non-amps switching modules
NOTE
Presently, non-amps Keithley modules include the Models 7701, 7703, 7706,
7707, 7708, and 7709. You can check the Keithley website (www.keithley.com)
for new modules.
A non-amp module does not support amps measurements. System channel operation can-
not be used to close channels while an amps function (DCI or ICI) is selected.
If an amps function (DCI or ACI) is selected and you attempt to close a system channel,
the message “NO AMPS CHAN” will be displayed briefly. For remote programming,
error -222 (Parameter data out of range) is generated. Example:
SYST:PRES
' Restores system preset defaults.
SENS:FUNC ‘CURR:DC’
' Selects DCI function.
ROUT:CLOS (@101)
' Attempts to close system channel 101 – Generates
' error -222.
If a system channel is already closed and you attempt to select the DCI or ACI function,
the message “INVALID FUNC” will be displayed briefly. For remote programming,
error -221 (Settings conflict) is generated. Example:
SYST:PRES
' Restores system preset defaults.
ROUT:CLOS (@101)
' Close system channel 101.
SENS:FUNC ‘CURR:DC’
' Attempts to select DCI function – Generates
' error -221.
2701-900-01.book Page 14 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...