B-4
Calibration Reference
Model 2701 Service Manual
Miscellaneous calibration commands
Miscellaneous commands perform miscellaneous calibration functions such as program-
ming the calibration code and date. These commands are discussed in detail in the following
paragraphs.
:CODE
(:CALibration:PROTected:CODE)
Purpose
To program the calibration code or password so that you can perform the
calibration procedures.
Format
:cal:prot:code '<char_string>'
Parameter
Up to a 8-character string including letters and numbers.
Description
The
:CODE
command enables the Model 2701 calibration procedures when
performing these procedures over the bus. In general, this command must be
sent to the unit before sending any other comprehensive or manufacturing
calibration command. The default calibration code is KI002701.
NOTES
The
:CODE
command should be sent only once before performing either the compre-
hensive or factory calibration. Do not send
:CODE
before each calibration step.
To change the code, first send the present code, then send the new code.
The code parameter must be enclosed in single quotes.
Example
:CAL:PROT:CODE 'KI002701'
Send default code of KI002701.
:COUNt?
(:CALibration:PROTected:COUNt?)
Purpose
To determine how many times the Model 2701 has been calibrated.
Format
:cal:prot:coun?
Response
<n> Calibration count.
Description
The
:COUNt?
command allows you to determine how many times the
Model 2701 has been calibrated.
NOTE
Use the
:COUNt?
command to help you monitor for unauthorized calibration
procedures.
Example
:CAL:PROT:COUN?
Request calibration count.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...