B-6
Calibration Reference
Model 2701 Service Manual
:LOCK?
(:CALibration:PROTected:LOCK?)
Purpose
To read comprehensive calibration lock status.
Format
:cal:prot:lock?
Response
0
Comprehensive calibration locked.
1
Comprehensive calibration unlocked.
Description
The
:LOCK?
query requests status from the Model 2701 on calibration
locked/unlocked state. Calibration must be enabled sending the
:CODE
command before calibration can be performed.
Example
:CAL:PROT:LOCK?
Request cal lock state.
:SAVE
(:CALibration:PROTected:SAVE)
Purpose
To save calibration constants in EEROM after the calibration procedure.
Format
:cal:prot:save
Parameter
None
Description
The
:SAVE
command stores internally calculated calibration constants
derived during both comprehensive and manufacturing calibration in
EEROM. (EEROM is non-volatile memory.) Calibration constants will be
retained indefinitely once saved. Generally,
:SAVE
is sent after all other
calibration steps (except for
:LOCK
).
NOTE
Calibration will be only temporary unless the
:SAVE
command is sent to
permanently store calibration constants.
Example
:CAL:PROT:SAVE
Save calibration constants.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...