B-10
Calibration Reference
Model 2701 Service Manual
:STEP2
(:CALibration:PROTected:DC:STEP2)
Purpose
To perform front terminal open-circuit calibration.
Format
:cal:prot:dc:step2
Parameter
None
Description
:STEP2
performs the open-circuit calibration step in the comprehensive
calibration procedure. Disconnect all cables and accessories from the input
jacks before sending this command.
Example
:CAL:PROT:DC:STEP2
Perform open circuit calibration.
:STEP3
(:CALibration:PROTected:DC:STEP3)
Purpose
To program the +10V comprehensive calibration step.
Format
:cal:prot:dc:step3 <Cal_voltage>
Parameter
<Cal_voltage> = 9 to 11 [V]
Description
:STEP3
programs the +10V DC comprehensive calibration step. The
allowable range of the calibration voltage parameter is from 9 to 11, but 10
is recommended for best results.
Example
:CAL:PROT:DC:STEP3 10
Program 10V step.
:STEP4
(:CALibration:PROTected:DC:STEP4)
Purpose
To program the -10V DC comprehensive calibration step.
Format
:cal:prot:dc:step4 <Cal_voltage>
Parameter
<Cal_voltage> = -9 to -11 [V]
Description
:STEP4
programs the -10V DC comprehensive calibration step. The allow-
able range of the calibration voltage parameter is from -9 to -11, but -10 is
recommended for best results.
Example
:CAL:PROT:DC:STEP4 -10
Program -10V step.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...