B-18
Calibration Reference
Model 2701 Service Manual
:CODE
(:CALibration:PROTected:CODE)
Purpose
To program the calibration code or password so that you can perform the
Model 7700 calibration procedures.
Format
:cal:prot:code '<char_string>'
Parameter
Up to a 8-character string including letters and numbers.
Description
The
:CODE
command enables the Model 2701 calibration procedures when
performing these procedures over the bus. This command must be sent to
the unit before sending any other Model 7700 calibration command. The
default calibration code is KI002701.
NOTES
The
:CODE
command should be sent only once before performing calibration. Do
not send
:CODE
before each calibration step.
The code parameter must be enclosed in single quotes.
Example
:CAL:PROT:CODE 'KI002701'
Send default code of KI002701.
:COUNt?
(:CALibration:PROTected:CARD1:COUNt?)
(:CALibration:PROTected:CARD2:COUNt?)
Purpose
To determine how many times a Model 7700 has been calibrated.
Format
:cal:prot:card1:coun?
:cal:prot:card2:coun?
Response
<n> Calibration count.
Description
The
:CARD1:COUNt?
and
:CARD2:COUNt?
commands allow you to
determine how many times a Model 7700 in Slot 1 and Slot 2 respectively
has been calibrated.
NOTE
Use the
:COUNt?
command to help you monitor for unauthorized calibration
procedures.
Example
:CAL:PROT:CARD1:COUN?
Request card 1 calibration count.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...