Model 2701 Service Manual
Calibration Reference
B-19
:DATE?
(:CALibration:PROTected:CARD1:DATE?)
(:CALibration:PROTected:CARD2:DATE?)
Purpose
To request the Model 7700 calibration date.
Format
:cal:prot:card1:date?
:cal:prot:card2:date?
Response
<year>, <month>, <day>
Description
The
:CARD1:DATE?
and
:CARD2:DATE?
queries allow you to read back
the calibration date from a Model 7700 in Slot 1 and Slot 2, respectively.
NOTE
The card calibration date is automatically set to the Model 2701 real time clock date
when the card is calibrated.
Example
:CAL:PROT:CARD1:DATE?
Request card 1 cal date.
:INIT
(:CALibration:PROTected:CARD1:INITiate)
Purpose
To initiate Model 7700 calibration procedures.
Format
:cal:prot:card1:init
Parameter
None
Description
The
:INIT
command enables Model 7700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the
:CODE
command, but before performing Model 7700
calibration.
Example
:CAL:PROT:CARD1:INIT
Initiate 7700 calibration.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...