1-20
Performance Verification
Model 2701 Service Manual
Verifying AC voltage
Check AC voltage accuracy by applying accurate AC voltages at specific frequencies from
the AC voltage calibrator to the Model 7700 inputs and verifying that the displayed readings
fall within specified ranges.
CAUTION
Do not exceed 300V RMS between plug-in module INPUT H and L
terminals or between adjacent channels, or 8
×
10
7
V•Hz input, because
instrument damage may occur.
Follow these steps to verify AC voltage accuracy:
1.
Connect the Model 7700 CH1 H and L INPUT terminals to the AC voltage calibrator as
shown in
Figure 1-9
.
Figure 1-9
Connections for Model 7700 AC volts verification
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
CH1
Output HI
Output
LO
Calibrator (Output AC Voltage)
Model 7700
Amplifier (Connect to calibrator)
Shielded
Cable
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...