Model 2701 Service Manual
Performance Verification
1-23
2.
Install the Model 7700 in Slot 1 of the Model 2701, then turn on the power and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3.
Select the DC current measurement function by pressing the
DCI
key.
4.
Set the Model 2701 for the 20mA range. Close Channel 21 by pressing the CLOSE key
and keying in 121.
5.
Source positive and negative full-scale currents for each of the ranges listed in
Table 1-11
, and verify that the readings for each range are within stated limits.
6.
Press the OPEN key to open Channel 21.
Table 1-11
Plug-in module DCI limits
DCI range
Applied DC current*
Reading limits (1 year, 18°C to 28°C)
20mA
100mA
1A
3A
20.0000mA
100.0000mA
1.000000A
3.000000A**
19.98840 to 20.01160mA
99.8700 to 100.1300mA
0.999120 to 1.000880A
2.99628 to 3.00372A
* Source positive and negative currents with values shown.
** If the Fluke 5725 amplifier is not available, apply 2.2A from calibrator. Reading limits for 2.2A
input are: 2.197240 to 2.202760A.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...