Model 2701 Service Manual
Performance Verification
1-29
Figure 1-14
Connections for Model 7700 thermocouple temperature verification
5.
Source each of the voltages summarized in
Table 1-14
and verify that the temperature
readings are within limits. Be sure to select the appropriate thermocouple type for each
group of readings. (See step 3 above.) Open Channel 1 after the test is complete.
Table 1-14
Model 7700 thermocouple temperature verification reading limits
Thermocouple type
Applied DC voltage*
Reading limits (1 year, 18°C to 28°C)
J
K
-7.659mV
0mV
42.280mV
-5.730mV
0mV
54.138mV
-191.0° to -189.0°C
-1.0° to +1.0°C
749.0° to 751.0°C
-191.0° to -189.0°C
-1.0° to +1.0°C
1349.0° to 1351.0°C
*Voltages shown are based on ITS-90 standard.
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
CH1
Make HI and LO
Connections
in Ice Bath
Ice Bath
Twisted
Thermocouple Wire
Output
HI
Output
LO
Calibrator (Output DC Voltage)
Notes: This setup and reading limits table
does not include errors from
ice point, thermocouple wire,
and connections.
HI and LO connections from
the calibrator and Model 7700
must be electrically isolated
from each other.
Model 7700
Low Thermal
Copper Connection
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...