Model 2701 Service Manual
Performance Verification
1-31
Verifying frequency
Follow the steps below to verify the Model 2701 frequency function:
1.
Connect the function generator to the Model 7700 CH1 H and L INPUT terminals. (See
Figure 1-15
.)
2.
Install the Model 7700 in Slot 1 of the Model 2701, then turn on the power and allow
the unit to warm up for one hour before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3.
Set the function generator to output a 1kHz, 1V RMS sine wave.
4.
Select the Model 2701 frequency function by pressing the FREQ key. Close Channel 1
by pressing the CLOSE key and keying in 101.
5.
Verify that the Model 2701 frequency reading is between 0.9999kHz and 1.0001kHz.
Figure 1-15
Connections for Model 7700 frequency verification
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
CH1
50
Ω
Coax Cable
Model 7700
Function Generator
Function
Output
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...