4-10
Troubleshooting
Model 2701 Service Manual
Analog circuitry
Refer to
Figure 4-3
for the following discussion on analog circuitry.
INPUT HI
INPUT HI protection is provided by the SSP (Solid State Protection) circuit. The SSP is
primarily made up of Q101 and Q102. An overload condition opens Q101 and Q102, which
disconnects the analog input signal from the rest of the analog circuit.
Note that for the 100VDC and 1000VDC ranges, Q101 and Q102 of the SSP are open. The
DC voltage signal is routed through the DCV Divider (Q114 and Q136) to the DCV switching
circuit.
AMPS input
The ACA or DCA input signal is applied to the Current Shunt circuit, which is made up of
K103, R158, R205, and R338. For the 20mA DC range, 5.1
Ω
(R205/[R338 + R158]) is
shunted across the input. Relay K103 is energized (set state) to select the shunts. For all other
DCA ranges and all ACA ranges, 0.1
Ω
(R158) is shunted across the input (K103 reset).
The ACA signal is then sent to the AC Switching & Gain circuit, while the DCA signal is
routed directly to the A/D MUX & Gain circuit.
Signal switching
Signal switching for DCV and OHMS is done by the DCV & Ohms Switching circuit. FETs
Q113, Q105, Q104, and Q108 connect the DCV or ohms signal to the X1 buffer (U113).
Note that the reference current for OHMS is generated by the Ohms I-Source circuit. For
4-wire ohms measurements, SENSE LO is connected to U126.
Signal switching and gain for ACV, FREQ and ACA is done by the AC Switching & Gain
circuit, which is primarily made up of K102, U102, U103, U105, U112, U118, U111, and
U110. Note that U111 is used for frequency adjustment. The states of these analog switches
vary from unit to unit.
Multiplexer and A/D converter
All input signals, except FREQ, are routed to the A/D MUX & Gain circuit. The multiplexer
(U163) switches the various signals for measurement. In addition to the input signal, the multi-
plexer also switches among reference and zero signals at various phases of the measurement
cycle.
When the input signal is selected by the MUX, it is amplified by U132 and U166. Gain is
controlled by switches in U129 and associated resistors.
The multiplexed signals of the measurement cycle are routed to the A/D Converter (U165)
where it converts the analog signals to digital form. The digital signals are then routed through
an opto-isolator to the MPU to calculate a reading.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...