3-18
Basic DMM Operation
Model 2701 User’s Manual
Model 7700 switching module
Connections for the Model 7700 switching module are shown in Figure 3-7. Note that
only channels 21 and 22 can be used for current measurements.
Figure 3-7
DCI and ACI connections using Model 7700 switching module
Amps measurement procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it has to be in the “F” (out) position. For switching modules, it has to be
in the “R” (in) position.
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Select the amps measurement function by pressing DCI or ACI.
3.
Use the RANGE and keys to select a measurement range consistent with the
expected current or press AUTO to select autoranging (AUTO annunciator turns
on). Details on range are provided in Section 4.
4.
Apply the current(s) to be measured.
CAUTION
Do not apply more than 3A to the input or the AMPS fuse will blow.
Model 7700 switching module — When performing current measure-
ments, the maximum allowable voltage is 60VDC or 30V RMS. Exceed-
ing these limits could cause damage to the switching module.
5.
If using a switching module, use the and keys to close the desired amps
channel (for the Model 7700, 21 or 22). All other channels will be open.
Model 7700
Switching
Module
H
L
CH 21 or 22
Caution: Maximum input: 60VDC or 30V RMS, 3A switched,
60W, 125VA maximum.
Current
Source
2701-900-01.book Page 18 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...