Model 2701 User’s Manual
Basic DMM Operation
3-27
Figure 3-10
Constant-current method to measure ohms (100
Ω
to 1M
Ω
ranges)
DUT
V
I
SOUR
V
MEAS
Input Hi
Input Lo
2701
DUT
V
I
SOUR
V
MEAS
Input Hi
Input Lo
2701
Sense Hi
Sense Lo
A) 2-wire ohms (
Ω
2) measurements (100
Ω
through 1M
Ω
ranges)
R
DUT
=
V
MEAS
I
SOUR
R
DUT
=
V
MEAS
I
SOUR
B) 4-wire ohms (
Ω
4) measurements (100
Ω
through 1M
Ω
ranges)
2701-900-01.book Page 27 Wednesday, August 3, 2011 9:43 AM