3-30
Basic DMM Operation
Model 2701 User’s Manual
Effects of open test leads on ohms readings
The Model 2701 will display readings up to 120% of range. Readings above 120% of
range will cause the “OVRFLW” message to be displayed. For example, on the 100
Ω
range readings up to 120
Ω
will be displayed. Above 120
Ω
, the “OVRFLW” message is
displayed.
The Model 2701 will also display the “OVRFLW” message if a test lead is open during an
ohms measurement. A hardware (H/W) detection circuit or software (S/W) detection is
used to detect an open input lead. For an
Ω
4 measurement, a software (S/W) detection rou-
tine is used to detect an open sense lead.
Open test lead detection is illustrated in Figure 3-12 for an
Ω
4 measurement of a 100
Ω
resistor using the 100
Ω
range. For an
Ω
2 measurement, sense circuity is not used. With the
test leads properly connected, as shown in Figure 3-12A, 1mA is sourced through the
100
Ω
DUT. The 100mV drop across the DUT appears on the Input Hi terminal. Resistance
is then calculated (100mV / 1mA = 100
Ω
) and displayed by the Model 2701.
Open input lead detection
100
Ω
through 1M
Ω
ranges
– For the lower ohms ranges, a hardware detector is used to
detect an ope n input lead. The hardware detector uses a comparator circuit to monitor the
voltage on the Input Hi terminal. Open circuit voltage on the Input Hi terminal is either
6.6V or 12.8V, depending on the selected measurement range (see resistance
specifications in Appendix A). When an input lead (Hi or Lo) is open, as shown in
Figure 3-12B, voltage rises to the open-circuit level, which trips the “OVRFLW” message.
10M
Ω
and 100M
Ω
ranges – For the two highest ohms ranges, open input lead detection is
implemented in software. Open circuit voltage for the 10M
Ω
and 100M
Ω
ranges is 7V. For
the 10M
Ω
range, the “OVRFLW” message trips when the open circuit voltage rises to
approximately 3.5V. For the 100M
Ω
range, the “OVRFLW” message trips when the open
circuit voltage rises to approximately 6.5V.
Open sense lead detection
100
Ω
through 1M
Ω
ranges – For the
Ω
4 function, the sense leads must be connected to the
DUT. As shown in Figure 3-10B, the sense leads connect the voltmeter of the Model 2701
to the DUT. In general, if a test lead for the voltmeter is open, the reading on the Model
2701 will randomly drift due to the high impedance circuitry of the voltmeter. If this were
allowed to happen for the
Ω
4 function, erroneous ohms readings would be displayed.
To prevent erroneous ohms readings caused by an open sense lead, the Model 2701
implements software to detect an open sense lead. As shown in Figure 3-12A, with all test
leads properly connected, voltage on Sense Hi is at virtually the same potential as Input
Hi, and Sense Lo is at virtually 0V. When a sense lead (Hi or Lo) opens, that terminal will
drift to -15mV and it will trip the “OVRFLW” message. Figure 3-12C shows detection for
an open Sense Hi lead.
2701-900-01.book Page 30 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...