3-44
Basic DMM Operation
Model 2701 User’s Manual
4-wire RTD temperature measurement configuration
The Alpha, Beta, Delta, and
Ω
at 0°C parameters for the five basic RTD types are provided
in Table 3-5. Note that these parameters can be modified using remote programming.
The steps to configure 4-wire RTD measurements are provided in Table 3-6. After
pressing SHIFT and then SENSOR, the menu starts at step 1 to select measurement units.
Each time you press ENTER to make a selection, the menu will automatically go to the
next selection. After pressing ENTER for the last step, the instrument will return to the
normal measurement state.
NOTE
As shown in Table 3-6, you can select the USER sensor type, but you cannot
change the USER parameters from the front panel. The parameters for the
USER type can only be set using remote programming (see TEMPerature:FRTD
commands in Table 3-7).
Table 3-5
RTD parameters
Type
Standard
Alpha
Beta
Delta
Ω
at 0°C
PT100
ITS-90
0.00385055
0.10863
1.49990
100
Ω
D100
ITS-90
0.003920
0.10630
1.49710
100
Ω
F100
ITS-90
0.003900
0.11000
1.49589
100
Ω
PT385
IPTS-68
0.003850
0.11100
1.50700
100
Ω
PT3916
IPTS-68
0.003916
0.11600
1.50594
100
Ω
Table 3-6
4-wire RTD temperature measurement configuration
Step
Menu Structure
Description
1
UNITS: C, F, or K
Select temperature measurement units
(°C, °F, or K).
2
SENS: 4W-RTD
Select the 4-wire RTD transducer.
3
TYPE: PT100, D100, F100,
PT385, PT3916, or USER
Select 4-wire RTD type.
2701-900-01.book Page 44 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...