Temperature measurement procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Select the temperature measurement function by pressing TEMP.
3.
Configure the temperature measurement as previously explained in “Temperature
measurement configuration,” page 3-42.”
4.
Connect the temperature transducer(s) to be measured.
5.
If using a switching module, perform the following steps to close the desired
channel. Keep in mind that for 4-wire RTD measurements, you will close the
primary (INPUT) channel (1 through 10). The channel that it is paired to will close
automatically.
a.
Press the CLOSE key.
b.
Use
,
, , and to key in the channel number and press ENTER. The
previously closed channel(s) (if any) will open and the specified channel (or
channel pair) will close.
NOTE
While in the normal measurement state, you can use the and keys to close
channels. In general, each key press will open the presently closed channel and
then close the next higher or lower channel.
6.
Observe the displayed reading.
7.
To measure other switching channels, repeat steps 5 and 6.
8.
When finished, press OPEN if there is a channel closed.
2701-900-01.book Page 45 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...