Model 2701 User’s Manual
Basic DMM Operation
3-47
Connections
NOTE
When using the front panel inputs, the INPUTS switch must be in the “F” (out)
position. For switching modules, it must be in the “R” (in) position.
Front panel input
When using the front panel input terminals, connect the test leads to the INPUT HI and
LO terminals as shown in Figure 3-17.
Figure 3-17
FREQ and PERIOD connections for front panel inputs
Model 7700 switching module
Connections for the Model 7700 switching module are shown in Figure 3-18. For this
2-wire measurement, channels 1 through 20 can be used.
Figure 3-18
FREQ and PERIOD connections using Model 7700 switching module
Input Impedance = 1M
Ω
in parallel with <100pF
Caution: Maximum Input = 1000V peak, 8 x 10
7
V
•
Hz
Model 2701
AC Voltage
Source
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
R
CAT I
H
L
CH 1-20
Model 7700
Switching
Module
AC
Voltage
Source
Caution: Maximum = 300V peak or RMS, 8 x 10
7
V
•
Hz
2701-900-01.book Page 47 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...