Model 2701 User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-19
Figure 4-4
Filter configuration flow chart
Scanning
The moving filter cannot be used when scanning. A scan channel cannot be configured to
use the moving filter. Also, the filter window is not used when scanning.
When a simple scan is configured, the present filter count and state will apply to all
channels in the scan. The window setting is ignored (effectively set to NONE) and if the
moving filter is selected, the filter will not enable when the scan is run.
For the advanced scan, filter state (on or off) and count can be set for each channel. You
cannot set unique filter count, type, and window settings from the advanced scan setup
menu.
For remote programming, the <clist> parameter is used to set filter count and state for
each channel in the scan. You cannot set unique filter type and window settings.
NOTE
Details on configuring a scan using filtering in a scan are provided in Section 7.
SHIFT
TYPE
WINDOW
0.01%
0.1%
1%
NONE
RDGS
001 to 100
TYPE
REPEAT
MOVNG AV
2701-900-01.book Page 19 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...