Model 2701 User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-3
3.
If using a switching module, use the or key to select (close) the input channel.
If using the front panel inputs (FRONT inputs selected), it does not matter if a
switching channel is closed.
4.
Press the REL key to set the rel value. The display will zero and the REL
annunciator will turn on.
5.
Apply the signal to be measured.
Pressing REL a second time disables rel.
You can input a rel value manually using the mX+b function. Set M for 1 and B for any
value you want. The mX+b function is covered in this section (see “Math,” page 5-7).
Scanning
When a simple scan is configured, the present rel setting will apply to all channels in the
scan. When an advanced scan is configured, each channel can have its own unique rel
setting. Details to configure and run a scan are provided in Section 7.
For an advanced scan, the following general procedure shows how to configure a scan
channel to use rel:
1.
While in the normal measurement state, select the appropriate function and close
the appropriate channel. For example, if scan channel 101 is going to be configured
for DCV and use rel, select DCV and close channel 101.
2.
Apply the DCV signal to be rel’ed to the closed channel. This could be an offset or
a baseline level.
3.
Press the REL key to enable rel (REL annunciator on). The input signal level is
used as the rel value.
4.
When configuring the advanced scan (as explained in Section 7), select the desired
channel, press DCV and then press REL (REL annunciator on).
When the channel is scanned, rel will enable using the rel value established in step 3.
For remote programming, the <clist> parameter is used to configure channels for a scan.
2701-900-01.book Page 3 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...