5-20
Rel, Math, Ratio, Channel Average, dB
Model 2701 User’s Manual
dB
Expressing DC or AC voltage in dB makes it possible to compress a large range of
measurements into a much smaller scope. The relationship between dB and voltage is
defined by the following equation:
where: V
IN
is the DC or AC input signal.
V
REF
is the specified voltage reference level.
The instrument will read 0dB when the reference voltage level is applied to the input.
If a relative value is in effect when dB is selected, the value is converted to dB then REL is
applied to dB. If REL is applied after dB has been selected, dB has REL applied to it.
NOTE
The dB calculation takes the absolute value of the ratio V
IN
/ V
REF
.
The largest negative value of dB is -160dB. This will accommodate a ratio of
V
IN
= 1µV and V
REF
= 1000V.
dB configuration
Remote programming must be used to configure the Model 2701 for dB measurements. It
cannot be configured from the front panel.
Scanning
Typically a scan using dB is configured and run using remote programming. However,
once dB is selected using remote programming, a simple dB scan can be configured and
run from the front panel. When the simple scan is configured, it will use the dB
measurement setup for each channel in the scan. Details on configuring and running a scan
are provided in Section 7.
NOTE
See
to configure and run a scan.
Remote programming — dB
dB commands
The dB commands are listed in Table 5-4. Details on these commands follow the table.
dB
20log
V
IN
V
REF
-------------
=
2701-900-01.book Page 20 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...