Model 2701 User’s Manual
Getting Started
1-9
Model 7709
Model 7710
Models 7711
and 7712
2-pole Operation 8-channels
20 channels
N/A
4-pole Operation 4 channel pairs
10 channel pairs
N/A
1-pole Operation N/A
N/A
8 channels
Measure Volts
300V maximum
60V maximum
No
3
Measure Amps
No
No
No
3
Measure Ohms
2/4-wire
2/4-wire
No
3
Thermocouple
Cold Junction
No
Yes
No
3
Relay Type
1
Latching electro-
mechanical
Solid state opto-cou-
pled FET
High frequency
electromechanical
Connector type
1 female DB-50
1 female DB-25
3.5mm removable
screw terminals
10 SMA
Configuration
2
6 x 8 matrix
Multiplexer
Multiplexer
Unique features
Rows 1 & 2 connect
to DMM (system
channel operation)
High-speed switch-
ing and long-life
relays
50
Ω
RF dual 1 x 4
multiplexer
Max Frequency:
7711: 2GHz
7712: 3.5GHz
1. Latching relays hold their open/close state after the mainframe is turned off. When turned on, all relays open after a few
seconds.
2. All multiplexers can be configured as two independent multiplexers.
3. The Models 7711 and 7712 have no measurement capabilities.
Table 1-1 (continued)
Model 77xx series switching modules
2701-900-01.book Page 9 Wednesday, August 3, 2011 9:43 AM
Summary of Contents for 2701
Page 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Page 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...